6. Fitting: Type of instrument or fitting to be verified. In Test box mode, select from BTE, ITE, ITC, CIC,
RITE (receiver in the ear), Body or FM. In On-ear mode, select from BTE, ITE, ITC, CIC, Open, Body or
FM. In the case of a CIC a deep insertion factor is applied to the estimated ear canal SPL.
7. Rule: Fitting rule. Select from DSL Adult, DSL Child, NAL-NL1, NAL-NL2 or None.
8. Type: Select from Binaural or Monaural fitting targets if DSL Adult, NAL-NL1 or NAL-NL2 has been
selected. For DSL child, only targets for a monaural fitting are provided.
9. Loss simulator: An electro-acoustic simulation intended to allow normal-hearing listeners to hear sounds as
if they had the elevated threshold and altered loudness perception caused by outer and inner hair cell
damage. See Sensory loss simulator.
Data entry
Threshold data and other audiometric parameters may be entered by scanning the barcode(s) on a previous
Audioscan printout or manually using a mouse or an external keyboard. See Mouse, keyboard, barcode scanner,
and Input device operation.
To enter data from a barcode see Barcode data input and:
1. Press the barcode trigger and center the red line on each barcode from a distance of 6 8 in. (15 20 cm).
The scanner will beep and the Barcode Entry poster will open. A green check mark indicates which data
have been entered. It does not matter in which order the barcodes are scanned or if each is scanned more than
once.
2. Click on [Done] to accept the data checked on the poster.
To enter audiometric data manually:
1. On the Threshold, UCL or RECD entry screens, enter each data point using a mouse or an external keyboard.
Left click the mouse to enter a point. Click again to delete the point. On the keyboard, the arrow keys change
frequency and level and the ENTER key on the numeric keypad enters or deletes a point. The keyboard's
numeric keypad can also be used to type in data.
2. Click on to save the data and exit the entry screen. Click on to exit the entry screen without
saving the data.
Fitting to targets for soft speech
See Speechmap screen choices for setup and audiometric data entry. See Using Speechmap.
1. Follow the instructions in Speechmap screen choices to enter audiometric data and select the fitting rule.
2. Click on to select the first Speechmap test. The Test 1 box is outlined to indicate it is active.